Selasa, 09 Februari 2016

MITECH MCT200 Coating Thickness Gauge

MITECH MCT200 Coating Thickness Gauge

MITECH MCT200 Coating Thickness Gauge

    This instrument is portable coating thickness gauge. It can test the coating thickness swiftly and accurately without damage but also can test the thin film thickness. It can be used widely in the detecting fields as manufacturing industry, metal-processing industry, chemical industry, commodity inspection and so on. It is the professional and necessary instrument for material pr
PRODUCT

 Specifications
Measuring principle: Magnetic induction & Eddy current
Measuring range: (01250) μm, depends on probes. maximum 10mm for the probe F10.  Refer to Appendix.
Low range resolution: 0.1μm
Accuracy ±(3%Thickness+1) μm, depends on probes and conditions
Display4 digits LCD with EL backlight
Memory for up to 20 files (up to 50 values for each file) of stored values.
Unit system: Metric (μm)Imperial (mil)
Power source: Two “AA” size, 1.5 Volt alkaline batteries. 200 hours typical operating time (EL backlight off).
CommunicationUSB1.1
Dimensions125mm×67mm×31 mm
Weight340g

Features
With different external probes, the gauge can be applied to measuring thickness of non-magnetic coating on magnetic metal substrate, as well as non-conductive coating on non-magnetic metal substrate.
Five types of probes (F400, F1, F10, N1, CN02) are available for different application. F10 probe measures up to 10mm.
Two measuring modes: single or continuous, changeable.
Two calibration methods can be applied to the gauge; And the system error of the probe can be corrected with the basic probe calibration method.
Measured values and user information are shown on a large, easy-to-read LCD. A display back light ensures easy reading of screen data in poorly-lit conditions.
Measuring status indicator showing the measuring status.
The gauges’ user-friendly measuring system permits automatic storage of up to 1000 readings in one memory matrix for later statistical evaluation.
Battery information indicates the rest capacity of the battery.
Auto sleep and auto power off function to conserve battery life.
USB1.1 communication port. Optional software to process the memory data on the PC.
Compact aluminum casesuitable for use under poor working conditions

Measuring Principles
The gauge adopts two thickness-measuring methods: magnetic induction method and eddy current method.
insulating coatings on all non-ferrous metals and on austenitic stainless steels, e.g. paint, anodizing coatings, ceramics, etc. applied on aluminum, copper, zinc die-casting, brass, etc.

Working Conditions
Working temperature -10℃~+50℃;
Storage temperature-30℃~+60℃;
Relative humidity: ≤90;
The surrounding environment should avoid of vibration, strong magnetic field, corrosive medium and heavy dust.

Configuration

No.
Item
Quantity
Remarks
Standard
Config.
1
Main unit
1

2
Probe
1
F1 or N1
3
Calibration Foils
5 PCS

4
Zero plate
1
Iron or Aluminum
5
Instrument case
1

6
Manual
1

7
Alkaline Battery
2 PCS
AA size
8



Optional
Config.
9
Other type of probes


10
DataPro software
1
On CD
11
USB cable
1


CV. ANDALAN PRIMA SEJAHTERA
JL. Muchtar Tabrani
Perum. Bintang Metropol Blok A7No 16
Bekasi Utara 17122
Phone : 021-29253184, 29253187
Fax     : 021-29253186
Mobile : 0812-9077-3137
               maria@andalanprimasejahtera.com
Web : www.andalanprimasejahtera.com

ALAT UJI TIDAK RETAK: MITECH MT160 Ultrasonic Thickness Gauge

ALAT UJI TIDAK RETAK: MITECH MT160 Ultrasonic Thickness Gauge: MITECH MT160 Ultrasonic Thickness Gauge The model MT160 is a digital ultrasonic thickenss gauge. Based on the same operatinfg principl...

Senin, 08 Februari 2016

MITECH MT150 Ultrasonic Thickness Gauge

MITECH MT150 Ultrasonic Thickness Gauge

    Ultrasonic Thickness Gauges are designed to improve safety and ensure reliability of material subject to corrosion or erosion. Corrosion gauges with their dual-element transducers are the primary inspection tool used to measure the remaining walls of components such as pipelines, pressure vessels, storage tanks, and numerous other applications. Precision gauges use single-element transducers to provide the thinnest range and highest degree of accuracy possible. Common applications include metals and a variety of non-metals (i.e. glass, ceramic, plastic, etc.).

FEATURE
Capable of performing measurements on a wide range of material, including metals, plastic, ceramics, composites, epoxies, glass and other ultrasonic wave well-conductive materials.
Transducer models are available for special application, including for coarse grain material and high temperature applications.
Two-Point Calibration function.
Two work modes: Single point mode and Scan mode.
Coupling status indicator showing the coupling status.
Auto sleep and auto power off function to conserve battery life.
SPECIFICATION
Display4.5 digits LCD with EL backlight.
Measuring Range0.75300mm (in Steel).
Sound Velocity Range 1000~9999 m/s.
MT150 0.1mm 
Accuracy ±0.5%Thickness+0.1mm, depends on materials and conditions
Units: Metric/Imperial unit selectable.
Four measurements readings per second for single point measurement, and ten per second for Scan Mode.
Memory for up to 20 files (up to 99 values for each file) of stored values.
Power SourceTwo “AA” size, 1.5 Volt alkaline batteries. 100 hours typical operating time (EL backlight off).
Outline dimensions150×74×32 mm. 
Weight
245g

CONFIGURATION


No
Item
Quantity
Note
 Standard Configuration
1
Main body
1

2
Transducer
1
Model: N05/90°
3
Couplant Bottle
1

4
Instrument Case
1

5

Alkaline battery
2
 AA size
6

Operating Manual
1

7

Warranty Card
1
8



Optional Configuration
9
Transducer: N02

10
Transducer: N07

11
Transducer: HT5

12
Mini thermal printer
1

13
Print cable
1

14
DataPro for Thickness Gauge
1
For use on the PC
15
Communication Cable
1


 
APPENDIX A: TRANSDUCER SELECTION

Model
Freq
MHZ
Diam
mm
Measuring RangeLower limitDescription
N022.514
3.0mm~300.0mm( In Steel)
 40mm (in Gray Cast Iron HT200)
20for thick, highly attenuating, or highly scattering materials
N055101.2mm~230.0mm( In Steel)Φ20mm×3.0mmNormal Measurement
N05/90°
5101.2mm~230.0mm( In Steel)Φ20mm×3.0mmNormal Measurement
N0776
 0.75mm~80.0mm
 ( In Steel)
Φ15mm×2.0mmFor thin pipe wall or small curvature pipe wall measurement
HT55123~200mm ( In Steel)30For high temperature (lower than 300°C) measurement.

APPENDIX B: SOUND VELOCITY IN DIFFERENT MATERIALS
Material
Sound Velocity
In/us
m/s
Aluminum
0.250
6340-6400
Steel, common
0.233
5920
Steel, stainless
0.226
5740
Brass
0.173
4399
Copper
0.186
4720
Iron
0.233
5930
Cast Iron
0.173-0.229
4400-5820
Lead
0.094
2400
Nylon
0.105
2680
Silver
0.142
3607
Gold
0.128
3251
Zinc
0.164
4170
Titanium
0.236
5990
Tin
0.117
2960
Epoxy resin
0.100
2540
Ice
0.157
3988
Nickel
0.222
5639
Plexiglass
0.106
2692
Polystyrene
0.092
2337
Porcelain
0.230
5842
PVC
0.094
2388
Quartz glass
0.222
5639
Rubber, vulcanized
0.091
2311
Teflon
0.056
1422
Water
0.058
1473
CV. ANDALAN PRIMA SEJAHTERA
JL. Muchtar Tabrani
Perum. Bintang Metropol Blok A7No 16
Bekasi Utara 17122
Phone : 021-29253184, 29253187
Fax     : 021-29253186
Mobile : 0812-9077-3137
               maria@andalanprimasejahtera.com